Type |
Specification |
GLP features |
Early maintenance feedback (EMF), electronic records of |
Communications |
Controller-area network (CAN). GPIB (IEEE-448), RS232C, |
Safety features |
Leak detection and safe leak handling, low voltages in maintenance areas, error detection and display |
Injection range |
0.1 – 100 μl in 0.1 μl increments |
Replicate injections |
1 – 99 from one vial |
Precision |
Typically < 0.5 % RSD of peak areas from 5 – 100 μl, |
Minimum sample volume |
1 μl from 5 μl sample in 100 μl microvial, or 1 μl from 10 μl sample in 300 μl microvial |
Carryover |
Typically < 0.1 %, < 0.05 % with external needle cleaning |
Sample viscosity |
range 0.2 – 50 cp |
Replicate injections per vial |
1 – 99 |
Sample capacity |
100 × 2-ml vials in 1 tray |
Injection cycle time |
Typically 50 s depending on draw speed and injection volume |
Type |
Specification |
Hydraulic system |
Dual plunger in series pump with proprietary servo-controlled variable stroke drive, floating plungers and active inlet valve |
Setable flow range |
0.001 – 10 ml/min, in 0.001 ml/min increments |
Flow range |
0.2 – 10.0 ml/min |
Flow precision |
< 0.3 % RSD (typically 0.15 %), based on retention time, at 1 ml/min |
Pressure |
Operating range 0 – 40 MPa (0 – 400 bar, 0 – 5880 psi) up to 5 ml/min Operating range 0 – 20 MPa (0 – 200 bar, 0 – 2950 psi) up to 10 ml/min |
Pressure pulsation |
< 2 %amplitude (typically < 1 %), at 1 ml/min isopropanol, |
Compressibility compensation |
User-selectable, based on mobile phase compressibility |
Recommended pH |
1.0 – 12.5, solvents with pH > 2.3 should not contain acids which attack |
Range |
stainless steel |
Gradient formation |
Low pressure quaternary mixing/gradient capability using proprietary |
Composition Range |
0 – 95 % or 5 – 100 %, user selectable |
Composition Precision |
< 0.2 % RSD, at 0.2 and 1 ml/min |
Control and data evaluation |
Agilent ChemStation for LC |
Analog output |
For pressure monitoring, 2 mV/bar, one output |
Communications |
Controller-area network (CAN), GPIB, RS-232C, APG Remote: ready, start, stop and shut-down signals, LAN optional |
Type |
Specification |
Temperature range |
10 degrees below ambient to 80 °C |
Temperature stability |
± 0.15 °C |
Temperature accuracy |
± 0.8 °C |
Column capacity |
Three 30 cm |
Warm-up/cool-down |
5 minutes from ambient to 40 °C |
Time |
10 minutes from 40 – 20 °C |
Dead volume |
3 μl left heat exchanger |
Dimensions |
6 μl right heat exchanger |
(h × w × d) |
(5.5 × 16 × 17 inches) |
Weight |
10.2 kg (22.5 lbs) |
Communications |
Controller-area network (CAN), GPIB, |
Type |
Specification |
Detection type |
Double-beam photometer |
Light source |
Deuterium lamp |
Wavelength range |
190–600 nm |
Short term noise |
± 0.75 × 10-5 AU at 254 nm |
(ASTM) |
|
Drift |
3 × 10-4 AU/hr at 254 |
Linearity |
> 2 AU (5%) upper limit |
Wavelength accuracy |
± 1 nm Self-calibration with deuterium lines, |
|
verification with holmium oxide filter |
Band width |
6.5 nm typical |
Flow cells |
Standard: 14-μl volume, 10-mm |
|
cell path length and 40 bar |
|
(588 psi) pressure maximum |
|
High pressure: 14-μl volume, |
|
10-mm cell path length and |
|
400 bar (5880 psi) pressure |
|
maximum |
|
Micro: 1-μl volume, 5-mm cell |
|
path length and 40 bar (588 psi) |
|
pressure maximum |
|
Semimicro: 5-μl volume, 6-mm |
|
cell path length and 40 bar |
|
(588 psi) pressure maximum |
Control and data |
Agilent ChemStation for LC |
evaluation |
|
Analog outputs |
Recorder/integrator: 100 mV or |
|
1 V, output range 0.001 – 2 AU, |
|
one output |
Communications |
Controller-area network (CAN), |
|
GPIB, RS-232C, APG Remote: |
|
ready, start, stop and |
|
shut-down signals, LAN |
|
optional |
Type |
Specification |
Detection type |
1024-element photodiode array |
Light source |
Deuterium and tungsten lamps |
Wavelength range |
190 – 950 nm |
Short term noise (ASTM) |
± 1 × 10-5 AU at 254 and 750 nm |
|
|
Single and Multi-Wavelength Drift |
2 × 10-3 AU/hr at 254 nm |
|
|
|
|
Linear absorbance range |
> 2 AU (upper limit) |
|
|
Wavelength accuracy |
± 1 nm Self-calibration with deuterium lines, |
|
verification with holmium oxide filter |
Wavelength bunching |
1 – 400 nm Programmable in steps of 1 nm |
Slit width |
1, 2, 4 , 8, 16 nm Programmable slit |
Diode width |
< 1 nm |
Flow cells |
Standard: 13 μl volume, 10 mm cell path |
|
length and 120 bar (1760 psi) |
|
pressure maximum |
|
Semi-Micro: 5 μl volume, 6 mm cell |
|
path length and 120 bar (1760 psi) |
|
pressure maximum |
|
High pressure: 1.7 μl volume, 6 mm cell |
|
path length and 400 bar (5880 psi) |
|
pressure maximum |
|
500 nano: 0.5 μl volume, 10 mm cell |
|
path length and 50 bar (725 psi) |
|
pressure maximum |
Control and data evaluation |
Agilent ChemStation for LC |
Analog outputs |
Recorder/integrator: 100 mV or 1 V, |
|
output range 0.001 – 2 AU, two outputs |
Communications |
Controller-area network (CAN), GPIB, |
|
RS-232C, APG Remote: ready, start, |
|
stop and shut-down signals, LAN optional |
• Flow rates from 50 uL/min to 5 mL/min can be generated for use with 2.1 mm ID columns and larger.
• The autosampler has a maximum capacity of 120 vials (12×32, 2-mL) with programmable temperature control from 4 to 40°C.
• A heated column compartment provides temperatures from 5 degrees above ambient to 65°C.
• The detector is a photodiode array (model 2996) with a wavelength range of 190-800 nm and sensitivity settings from 0.0001-2.0000 absorbance units.
• Flow rates from 50 uL/min to 5 mL/min can be generated for use with 2.1 mm ID columns and larger.
• The autosampler has a maximum capacity of 120 vials (12×32, 2-mL) with programmable temperature control from 4 to 40°C.
• A heated column compartment provides temperatures from 5 degrees above ambient to 65°C.
• The detector is a photodiode array (model 2996) with a wavelength range of 190-800 nm and sensitivity settings from 0.0001-2.0000 absorbance units.
Refurbished TSP HPLC System comes with a TSP AS3000 Variable Volume Autosampler , TSP P1000 Isocratic Pump, and TSP UV 1000 Detector
The TSP AS3000 Autosampler provides outstanding
performance and maximum flexibility in an LC sample processing system. The
standard unit offers 120-vial capacity, multiple injection modes, random vial
access, re-use of calibration standards, and priority sampling
• Vial capacity:
– 120 vials in three removable trays
– 105 vials with tray temperature control
• Variable-Volume:
– Injection Precision: <0.5 % RSD @ 10 pL or greater
– Injection Volume: 0.1 – 100 pL injection standard (250 uL syringe) up to 1000 uL injections with larger loops and syringes
• Fixed-Loop:
– Injection Precision: <0.5% RSD @ 10 uL or greater
– Injection Volume: 20 uL standard (250 uL syringe) 1000 uL injections with larger loops and syringes
• Sample Carryover:
– <0.0 I% at 300 gL flush volume
– 5 mL flush volumes can be set
• Minimum Sample Volume: 1 uL can be injected from 5 uL with an end-port needle
• Needle Height: Programmable in 0.1 mm increments
• Method Files: Protected in non-volatile memory
• Communications
– Remote Inputs: Pump Ready, Inject Hold
– Outputs Pump Stop, Autosampler Ready, Inject, Gradient Start, Four timed-event outputs
– SpectraNETIRS-232 port
• Ambient Environment: 10° – 40°C, 5 – 95% relative humidity
• Dimensions: 37cm x 41 cmx 51 cm (HxWxD)
TSP UV 1000 Detector is a programmable variable single wavelength
detector. It has superior optical performance and has externally mounted
interchangeable flow cells for wide range of applications ranging from
capillary, microbore to preparative, CE, SFC and LC/MS.
Wavelength Ranges: D2 lamp: 190 to 380 nm, W lamp: 366 to 800 nm
• Wavelength Accuracy: ± 1.0 nm
• Wavelength Precision: <0.01 nm
• Bandwidth: 6 nm
• Range Selections: 3.0, 2.0, 1.0, 0.5, 0.2, 0.1, 0.05, 0.02, 0.01, 0.005, 0.002, 0.001, 0.0005 AUFS
• Absorbance Range: 0.0005 to 3.0 AUFS
• Absorance Linearity @ 254 nm: 3.0 AUFS
Communications:
– Remote Inputs: Run, Stop, and Zero
– Remote Output: Ready
- L-2130 HTA Pump
- L-2200 Autosampler
- L-2300 Column Oven
- L-2490 Refractive Index Detector
- L-2485 Fluorescence Detector
- L-2400 UV Detector
- L-2455 Diode Array Detector
L-2130 HTA Pump Specifications
|
|
Accuracy |
+/- 2 µL (0.01 to 0.10 mL/min) +/- 2 % (0.101 to 8.000 mL/min) +/- 4 % (8.001 to 9.999 mL/min) |
Precision |
+/-0.075 % RSD or 0.02 min SD at 1.0 mL/min |
Constant pressure |
290 to 5600 psi at 0 to 5.0 mL/min |
Solvent delivery |
100 µL stroke |
Dimensions |
13.5 in (W) x 15.75 in (D) x 6.0 in (D) (340 mm (W) x 400 mm (D) x 150 mm (D)) |
Weight approx. |
24 lbs (11 kg) |
HIGH PRESSURE GRADIENT
|
|
Flow rate range |
0.001 to 9.999 mL/min (0.2 to 5.0 mL/min recommended) |
Binary |
Up to2 eluents |
Composition range |
1 to 100% (0.1% increments) |
Accuracy |
+/- 1 % (0.2 mL/min) |
Precision |
0.1% RSD or 0.02 min SD at 1.0 mL/min |
LOW PRESSURE GRADIENT |
|
Flow rate range |
0.001 to 9.999 mL/min (0.2 to 5.0 mL recommended) |
Quarternary |
Up to 4 eluents |
Composition range |
0 to 100% (0.1% increments with CDS; 1% increments with front
panel control |
Accuracy |
+/- 1 % (0.2 mL/min) |
Precision |
0.1 % RSD or 0.02 min SD at 1.0 mL/min |
L-2200 Autosampler SpecificationsL |
|
Carry over |
0.01% (< 0.003% with typical results) |
Injection method |
Direct all injection |
Capacity |
200 vials (1.5 mL) |
Cooling options |
4 to20 °C (at ambient) |
Rack options |
128 (4 mL), 336 (1 mL), Microtiter rack (3 plates) |
Injection volume |
0.1 to 50 µL (standard), 50 µL to 4.5 mL (optional) |
Repeatability |
0.3% RSD (at 10-µL injections) |
% RSD of slopes |
Correlation coefficient > 0.999 Response factor < 5% RSD |
Max. # injections/vial |
99 |
Max. cycle time |
999.99 |
Min. cycle time |
30 s |
Syringe volumes |
0.100 mL (standard), 0.5, 1, 2.5, and 5 mL (optional) |
Safety |
Built-in leak sensor |
Dimensions |
13.5 in (W) x 15.75 in (D) x 12.0 in (D) (340 mm (W) x 400 mm (D) x300 mm (D)) |
Weight approx. |
42 lbs (19 kg) |
L-2300 Column Oven Specifications
|
|
Range |
[Room temp - 15 °C] to[Room temp + 50°C]
) |
Setting range |
1 to 65 °C |
Accuracy |
+/- 1 °C |
Safety |
Built-in vapor/leak sensor |
Stability |
+/- 0.1 °C |
Column capacity |
Up to three 25-cm columns |
GLP functions |
Serial number registration in unit memory |
Dimensions |
13.5 in (W) x 15.75 in (D) x 6.0 in (D) (340 mm (W) x 400 mm (D) x 150mm (D)) |
Weight approx. |
22 lbs (10 kg) |
L-2490 RI Detector Specifications
|
|
Method of Detection |
Deflection |
Light source |
Tungsten lamp |
Refractive index range |
1.0 to 1.75 |
Linearity range |
600µRIU |
Baseline noise |
2.5 x 10-9 RIU or less (water filled in, time constant 3.0
s) |
Temperature control |
Off, 30 °C to 50 °C |
Time constants |
0.1, 0.25, 0.5, 1, 1.5, 2, 3, 6 s |
External outputs |
Warning when measurement range is exceeded, Auto zero, Alarm,
Recorder 10 mV/FS, Integrator 1 V/FS |
External inputs |
Auto zero, Marker, Purge |
Baseline adjustment range |
0 to 50% /FS |
Auto zero |
Full auto zero (optics + calculation) |
Recorder range |
0.25 to 512 x 10-6 RIU/FS |
Cell volume |
8 µL |
Max. flow rate |
10 mL/min |
Max. pressure in cell |
7.1 psi (0.049 MPa) |
Dimensions |
13.5 in (W) x 15.75 in (D) x 6.0 in (D) (340 mm (W) x 400 mm (D) x 150 mm (D)) |
Weight Approx. |
29 lbs (13kg) |
Power requirement & consumption |
100 to 240 VAC (+/- 10%), 50/60 Hz, 150 VA or less |
L-2485 Fluorescence
Detector Specifications |
|
Optics |
Ratio-beam photometric detector using transmission
light |
Light source |
150-W Xe lamp Hg lamp for wavelength check |
EX wavelength range and bandwidth |
200 to 850 nm 15 nm |
EM wavelength range and bandwidth |
250 to 900 nm (over 731 nm requires PMT exchange) 15 nm and 30 nm |
Sensitivity |
S/N: >900 for Raman spectrum of water (EX: 350 nm; time constant: 2 sec; bandwidth:30 nm; baseline method) |
Dimensions |
13.5 in (W) x 15.75 in (D) x12.0 in (D) (340 mm (W) x 400 mm (D) x 300 mm (D)) |
Weight approx. |
42 lbs (19 kg) |
L-2400 UV/L-2420 UV/VisDetectors
Specifications |
|
Optics |
Double beam and ratio mode |
Calibration |
Built-in Hg lamp |
Range |
L-2400: 190 to 600 nm L-2420: 190 to 900 nm |
Light source |
L-2400: D2 lamp L-2420: D2 lamp and W lamp |
Accuracy |
+/- 1 nm |
Bandwidth |
6 nm |
Noise |
< 0.6 x 10-5 AU |
Drift |
< 1.0 x 10-4 AU/h |
AUFS range |
0.25, 0.5, 1, 2 AU |
Response |
0.05, 0.1, 0.5, 1, 2, 4, 8 s |
Programs |
9 programmable steps |
Max. Time |
600 min |
Max. Steps |
100 (in 9 programs) |
Available functions |
Wavelength, time, autozero, and hold |
GLP functions |
Auto calibration using Hg lamp Lamp energy check Log for usage Replacement date Serial number registration in unit memory |
Safety |
Bult-in leak sensor, Error handling |
Dimensions |
13.5 in (W) x 15.75 in (D) x 6.0 in (D) (340 mm (W) x 400 mm (D) x 150mm (D)) |
Weight approx. |
24 lbs (11 kg) |
L-2455 Diode Array Detector
Specifications |
|
Optics |
Single-beam photometric detector with grating
monochromator and 1024-bit diode array |
Light source |
D2 and W lamps Hg lamp for wavelength calibration |
Wavelength range |
190 - 900 nm |
Noise |
< 0.5 x 10-5 AU (at 250 nm and 600 nm) |
Drift |
< 0.5 x 10-3 AU/h (at 250 nm and 600 nm) |
Slit width |
1 nm or 4 nm |
Wavelength time program |
by EZChrom Elite |
Spectrum scan function with background subtraction |
by EZChrom Elite |
GLP functions |
Auto calibration using Hg lamp Lamp energy check Log for usage Lamp replacement date Serial number registration in unit memory |
Safety |
Leak sensor, Error handling |
Input/Output contact signals |
START IN BUSY OUT ERROR IN/OUT |
Power requirements |
24 V DC, 100 W |
Dimensions |
13.5 in (W) x 15.75 in (D) x 6.0 in (D) (340 mm (W) x 400 mm (D) x 150mm (D)) |
Weight approx. |
26 lbs (12 kg) |
Agilent 6890 |
Agilent 6890N |
|
---|---|---|
Height |
50 cm (19.7 in.) |
50 cm (19.7 in.) |
Width |
58 cm (26.8 in.) |
58 cm (26.8 in.) |
Depth |
54 cm (21.6 in.) |
54 cm (21.6 in.) |
Average Weight |
49 kg (108 lb) |
49 kg (108 lb) |
Networkable |
No |
Yes |
6890 Environmental Conditions |
||
Ambient Operating |
15 ˚C to 35 ˚C |
15 ˚C to 35 ˚C |
Ambient Operating |
5-95 % |
5-95 % |
Storage Extremes |
-40˚C to 65˚ |
-40˚C to 65˚ |
Line Voltage Requirements |
±5% of nominal |
±5% of nominal |
Agilent 5890 Column Oven |
||
Dimensions |
28 x 31 x 16 cm |
28 x 31 x 16 cm |
Operating temperature |
4˚C above ambient to 450˚C |
4˚C above ambient to 450˚C |
Maximum run time |
120˚C/min |
120˚C/min |
Available Detectors for the |
Flame ionization detector (FID) |
Flame ionization detector (FID) |
FID
- Maximum temperature: 450 oC
- Detectivity: 2 pg C/sec
- Linear dynamic range: 107
- Patented ceramic flame tip: standard
- Flame-out detection/auto re-ignition: standard
CP-1177 Split/Splitless
- Pressure range: 0-100 psi
- Total flow: 500 mL/min at 10 psi
- Maximum temperature: 450 oC
- Split range: 1-10,000 (column dependent)
Column Oven
- Dimensions: 28 cm (w) x 20 cm (d) x 28 cm (h)
- Temperature range: ambient +4 oC to 450 oC
- With liquid N2: -99 oC to 450 oC
- With liquid CO2: -65 oC to 450 oC
- Temperature program ramps/holds: 7/8
- Maximum temperature ramp rate: 100 oC /min for all voltages
- Cool down rate: 400 oC to 50 oC in 4.5 minutes
- Temperature set-point resolution: 1 oC
Dimensions and Weights
Agilent 6890 |
Agilent 6890N |
|
---|---|---|
Height |
50 cm (19.7 in.) |
50 cm (19.7 in.) |
Width |
58 cm (26.8 in.) |
58 cm (26.8 in.) |
Depth |
54 cm (21.6 in.) |
54 cm (21.6 in.) |
Average Weight |
49 kg (108 lb) |
49 kg (108 lb) |
Networkable |
No |
Yes |
6890 Environmental Conditions |
||
Ambient Operating |
15 ˚C to 35 ˚C |
15 ˚C to 35 ˚C |
Ambient Operating |
5-95 % |
5-95 % |
Storage Extremes |
-40˚C to 65˚ |
-40˚C to 65˚ |
Line Voltage Requirements |
±5% of nominal |
±5% of nominal |
Agilent 5890 Column Oven |
||
Dimensions |
28 x 31 x 16 cm |
28 x 31 x 16 cm |
Operating temperature |
4˚C above ambient to 450˚C |
4˚C above ambient to 450˚C |
Maximum run time |
120˚C/min |
120˚C/min |
Available Detectors for the |
Flame ionization detector (FID) |
Flame ionization detector (FID) |
FID
- Maximum temperature: 450 oC
- Detectivity: 2 pg C/sec
- Linear dynamic range: 107
- Patented ceramic flame tip: standard
- Flame-out detection/auto re-ignition: standard
CP-1177 Split/Splitless
- Pressure range: 0-100 psi
- Total flow: 500 mL/min at 10 psi
- Maximum temperature: 450 oC
- Split range: 1-10,000 (column dependent)
Column Oven
- Dimensions: 28 cm (w) x 20 cm (d) x 28 cm (h)
- Temperature range: ambient +4 oC to 450 oC
- With liquid N2: -99 oC to 450 oC
- With liquid CO2: -65 oC to 450 oC
- Temperature program ramps/holds: 7/8
- Maximum temperature ramp rate: 100 oC /min for all voltages
- Cool down rate: 400 oC to 50 oC in 4.5 minutes
- Temperature set-point resolution: 1 oC
Dimensions and Weights